Quick Field Instructions

For Federal Communications Commission

Proof Testing on the

Tektronix® 2714 Cable Television

Radio Frequency Analyzer

 

 A Condensation by George Stantial

 

   

Introduction

The typical spectrum analyzer in use for cable television purposes is a swept-tuned or scanning radio-frequency analyzer. This means that instead of displaying signal amplitude as a function of time as is done by the oscilloscope, the spectrum analyzer will use the frequency domain as the x-axis on its graphical display.   This allows radio energy signals differing in frequency to be examined for their proper attributes.  The technician using this field guide should already be aware of this basic information along with the uses and limitations of spectrum analysis.  The basic information presented in this quick field guide is not meant to substitute for the proper understanding and operation of the Tektronix® 2714 CATV Analyzer.  Indeed, the instructions assume the technician already has some familiarity with this analyzer.  Additionally, the technician should have a basic understanding of the tests required by the Federal Communications Commission (FCC).  The purpose of this field guide is to allow experienced technicians to quickly perform certain required measurements without the need to search and summarize the supplied manual.  This guide is neither supplied nor endorsed by the manufacturer of Tektronix® equipment. To use the Tektronix® analyzer for purposes other than those outlined below, it is recommended the Tektronix® provided manual be consulted.

The finger controls on the Tektronix® spectrum analyzer will allow the user to move and zoom (in terms of frequency and amplitude) on the displays of the received signals of interest. Other controls will allow certain operating parameters on the analyzers such as "bandwidth" and "scan rate" to be changed.  Properly setting the parameters will not only increase the accuracy of measurements, but may allow anomalies to be seen that could not be seen otherwise.  Yet more controls allow the software-driven manipulation of the previously mentioned controls to perform automated multi-step testing. 

A spectrum analyzer is a versatile testing and troubleshooting device whose accuracy limits must be respected. The user must be aware of the specifications of the analyzer and be able to determine if it meets the needs for its intended purpose.  As an example, many times, the frequency display portion of a spectrum analyzer is a reference only and not an accurate depiction of frequency. The analyzer must have a built-in frequency counter to read actual frequency.  For FCC testing, the counter must have a high enough accuracy for that purpose. The amplitude reading is also subject to accuracy ratings. Furthermore, some tests may require signal amplitudes to be within a specific range to reduce the uncertainty of the result of those tests. In all cases, refer to the owner’s manual for the specifications or consult with the manufacturer.

 

Setup Instructions for the Tektronix® 2714 CATV Analyzer

For the purposes of the instruction listed below, labels in capital letters are for menu items on the display and will be preceded by a “Select” command.  Labels in bold capital letters are for physical buttons on the analyzer and will be preceded by a “Press” command.

Press POWER

Allow the analyzer to warm up 15 minutes before use.

Due to the variety of older machines in use, the following instructions are based on analyzer firmware Version 06.05.92. The firmware should be upgraded when the analyzer is sent in for calibration.

To check the version of firmware in use perform the following steps:

Press UTIL in the button grouping labeledMENUS”

Select 4 SYSTEM CONFIGURATION

Select 9 INSTALLED OPTIONS DISPLAY

The version number (date) will read under the INSTALLED OPTIONS heading

Press any “MENUS” key to exit the utility menu

To check/modify the test setup perform the following steps:

Press CATV/APPL in the button grouping labeled “MENUS”

If the menu heading does not read CATV MEASUREMENTS, Select 9 to leave its current menu

Select 5 CARRIER TO NOISE to enter that menu (Select 9 MORE if required first)

Select 5 CARRIER TO NOISE SETUP to enter that menu

If needed, Select 5 NOISE NORM’D BW, Press 4 and Press MHz mSEC mV on the keypad for 4.0 MHZBW

Select 0 INTERACTIVE to place the C/N test in interactive mode

Do not select 1 AUTO as firmware older than 01.05.95 will violate FCC parameters

Do not select 2 AUTO (PAUSE FOR CARRIER OFF) as firmware older than 01.05.95 will violate FCC parameters

Press BSKP as needed to return to the CATV MEASUREMENTS menu

Select 6 HUM/LFD to enter that menu

If needed, Select 5 POWER LINE FREQ, to toggle the line frequency to 60Hz

Press BSKP as needed to return to the CATV MEASUREMENTS menu

Select 5 CTB/CSO to enter that menu (Select 9 MORE if required first)

Select 5 CTB/CSO SETUP to enter that menu

Select 0 INTERACTIVE if it is desired to manually select the beats to measure

Do not select 1 AUTO as CTB cannot be measured without carrier removal

Select 2 AUTO (PAUSE FOR CARRIER OFF) for a slow measurement with better accuracy

Select 3 SINGLE SWEEP for a speedy measurement with good accuracy

The AUTO TEST LOCATIONS should read:

4 +1.250 MHZ (OFFSET)

5 +750.000 MHZ (OFFSET)

6 + 0.000 MHZ (OFFSET)

7 -750.000 MHZ (OFFSET)

8 -1.250 MHZ (OFFSET)

Optionally, the lowest offset may be deleted if the upper band-edge of the lower adjacent channel creates false CSO

To delete it, Select 8 –1.250 MHZ (OFFSET), Press W=DELETE

Press BSKP as needed to return to the CATV MEASUREMENTS menu

Select 7 IN-CHANNEL RESPONSE to (Select 9 MORE if required first)

Select 5 IN-CHAN RESP SETUP to enter that menu

Select 0 INTERACTIVE for manual placement of the high/low markers on the headend injected multiburst pulses

Select 1 AUTO for the recommended automatic placement of the high/low markers on the multiburst pulses

Press BSKP as needed to return to the CATV MEASUREMENTS menu or any “MENUS” key to exit

Setup for these tests is now complete

 

Test Instructions for the Tektronix® 2714 CATV Analyzer

To Tune to the Channel Desired for Test:

Press channel number(s) desired on KEYPAD

Press W on KEYPAD

 

To Test Carrier to Noise:

Press CATV/APPL in the button grouping labeled “MENUS”

If the analyzer is in a test mode other than the one desired, Select 9 to leave

If needed, Select 5 CARRIER TO NOISE to enter that menu (Select 9 MORE if required first)

Select 0 RUN CARRIER/NOISE INTERACTIVE

For unmodulated carrier, on MOVE MARKER TO NOISE prompt move marker near channel center (+/- 2MHz)

Press W to continue

For modulated carrier, on MOVE MARKER TO NOISE prompt first remove carrier, then move marker as above

Press W to continue

Record the displayed results

Tune to the next channel desired and repeat the steps listed above

 

To Test CTB/CSO:

Tune to the channel desired

Press CATV/APPL in the button grouping labeled “MENUS”

If the analyzer is in a test mode other than the one desired, Select 9 to leave

If needed, Select 5 CTB/CSO to enter that menu (Select 9 MORE if required first)

Select 2 RUN CTB/CSO to run the test in the mode selected during setup

When prompted, turn off the carrier

Press W to continue and follow any further prompts depending on test mode

The display will show the worst case

Press CATV/APPL

Select 2 DISPLAY RESULTS

All results for both both CTB and CSO will display

Record the displayed results (only worst case for each of CSO and CTB is required)

Press W to continue

Press any “MENUS” key to exit

Tune to the next channel desired and repeat the steps listed above

 

To Test Hum and other Low Frequency Disturbances:

Tune to the channel desired

Press CATV/APPL in the button grouping labeled “MENUS”

If the analyzer is in a test mode other than the one desired, Select 9 to leave

If needed, Select 6 HUM/LFD to enter that menu (Select 9 MORE if required first)

Remove modulation (recommended) for a system Hum/LFD reading

Leave modulation for a reading of all LFD (source video may influence reading)

Select 0 RUN HUM/LFD

Record the displayed result

This test is only needed on one channel at each test location

 To Test In-Channel Response:

Tune to the channel desired

Press CATV/APPL in the button grouping labeled “MENUS”

If the analyzer is in a test mode other than the one desired, Select 9 to leave

If needed, Select 7 IN-CHANNEL RESPONSE to enter that menu (Select 9 MORE if required first)

Select 0 RUN IN-CHANNEL RESPONSE

Inject the multiburst test signal when prompted

Press W to continue

Remove the multiburst test signal when prompted

Press W to continue

Record the displayed result

Tune to the next channel desired and repeat the steps listed above

 

  Tektronix® is a registered trademark of Tektronix Inc.

 

Home

Copyright © 1998-2009 G. Stantial